| Model Number |
G148 |
| Medical Device Problem Codes |
Over-Sensing (1438); Under-Sensing (1661)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/16/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during a routine generator change procedure, the presenting electrogram (egm) of this cardiac resynchronization therapy defibrillator (crt-d) showed farfield atrial oversensing and undersensing of atrial flutter.Technical services (ts) was contacted and provided programming optimization recommendations.The generator change was completed successfully with the crt-d explanted due to normal battery depletion (nbd) and replaced with a new device and the right atrial (ra) lead remains in service.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|