| Model Number |
Patient Information Center iX |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/12/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on a patient information center ix which indicated wireless telemetry is not visible on the central monitor in several units hospital wide although hard-wired monitors are visible.The device was in use on a patient at time of event, there was no adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis was performed by onsite service personnel which included functional testing and troubleshooting.The results of the analysis confirmed the issue and identified that on the customer side; a switch was down.Based on the results of the analysis, it was determined that the product has malfunctioned and the most likely cause of the reported problem was a switch that was down on the customer side of the network.The issue was resolved by plugging the uninterrupted power supply (ups) back in.The system was checked after bringing it back up and found to be resolved.The product is back in service.Based on the information available and results of additional analysis, no further action is necessary at this time.If additional information is received, the complaint file will be reopened for further investigation.
|
| |
|