| Model Number |
DDPB3D4 |
| Medical Device Problem Codes |
Intermittent Capture (1080); High impedance (1291); Mechanical Problem (1384); Pacing Problem (1439)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/22/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) had exhibited connection issues as the setscrew was loose and caused pacing problems and no capture.An x-ray indicated that the right atrial (ra) lead pin was not properly set in the header.The device interrogation indicated that there was intermitted capture and out of range (oor) high atrial pacing impedance.The lead was removed from the header and tested with no noise noted.The icd was explanted and replace, the lead was reposition and re-implanted into the new device and remains in use. no patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|