• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION CORPORATION; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION CORPORATION; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS Back to Search Results
Model Number DB-2203-45
Medical Device Problem Codes High impedance (1291); Use of Device Problem (1670); Low impedance (2285)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/16/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a routine deep brain stimulation (dbs) lead placement procedure, a lead located in the left part of the brain, which had been implanted during a previous procedure, was found to present multiple low and high impedances.The doctor explored the connection between the left brain electrode and the extension wire and identified a physical issue, as the lead was positioned at a 90-degree angle at the junction where the lead plugs into the extension wire.The doctor stated that the lead must have been damaged during the initial implant.The doctor then proceeded to straighten the kinked left brain electrode; however, this had no effect on the abnormal impedance readings.As the routine procedure did not contemplate replacement of the damaged lead, the doctor decided not to replace it and instead perform the programming around those impedances.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN SYMPTOMS
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
Boston Scientific Corporation
road 698, lot no. 12
dorado PR 00646 -260
PR   00646-2602
Manufacturer Contact
ricardo amaro castro
25155 rye canyon loop
valencia, CA 91355
+1(320)800-7570
MDR Report Key50168157
Report Number3006630150-2026-05644
Device Sequence Number5305666
Product Code NHL
UDI-Device Identifier08714729966340
UDI-Public(01)08714729966340(17)271021(10)5005971
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberDB-2203-45
Device Catalogue NumberDB-2203-45
Device Lot Number5005971
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/14/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date08/06/2026
Supplement Report FDA Received Date09/03/2026
Date Device Manufactured10/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexMale
-
-