It was reported that during a routine deep brain stimulation (dbs) lead placement procedure, a lead located in the left part of the brain, which had been implanted during a previous procedure, was found to present multiple low and high impedances.The doctor explored the connection between the left brain electrode and the extension wire and identified a physical issue, as the lead was positioned at a 90-degree angle at the junction where the lead plugs into the extension wire.The doctor stated that the lead must have been damaged during the initial implant.The doctor then proceeded to straighten the kinked left brain electrode; however, this had no effect on the abnormal impedance readings.As the routine procedure did not contemplate replacement of the damaged lead, the doctor decided not to replace it and instead perform the programming around those impedances.
|