It was reported that the coil detached requiring intervention, and there were unretrieved device fragments.A 4mm x 15cm embold? fibered coil was selected for use in an embolization procedure to treat a bleed.The vessel was mildly tortuous.During the procedure, difficulty advancing was noted, and the coil detached and fell apart in the 130cm truselect microcatheter while the coil was half in the vasculature and half in the microcatheter.The proximal release perforation was not broken.The coil was deployed into the gastroduodenal artery (gda) while prolapsing into the common, proper and right hepatic arteries, which it subsequently thrombosed.Most of the coil was retrieved with a snare, but there was a stringy portion that was coming off the end of it and detached.Most of that portion was retrieved too, but it was too late for the common hepatic, proper hepatic, and gda, which were all completely thrombosed.The majority of the device fragments were removed during the procedure, and there is no plan to remove any fragments that remain.It was also noted that the procedure was aborted because of diffuse vessel thrombosis / dissection.The patient was in stable condition and discharged.
|
|
Detailed product information was not provided to boston scientific.We completed a good faith effort to obtain the information, but it was unable to be obtained.Because the lot # is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.
|