H3) the quick-cross was returned for evaluation.Visual inspection found the proximal marker band moved distally next to the middle band, which is approx.54 mm from its expected position.The middle marker band moved approx.4 mm distal to its expected position.The distal marker band is in the correct position.There was no evidence of elongation; however, the tube is pinched 30-50 mm from the distal tip.Original swagging marks are visible in their expected positions, and encapsulation is in good condition.H6) based on device evaluation and investigation, movement of the marker bands was confirmed; however, the cause of the movement could not be established.Type of investigation code b01 replaced (from b17).Investigation findings code c0706 replaced (from c20).All other codes are applicable as listed in the initial mdr.Submission of this report does not, in itself, represent a conclusion by the manufacturer and/or authorized representative or the national competent authority that the content of this report is complete or accurate, that the medical device(s) listed failed in any manner and/or that the medical device(s) caused or contributed to an alleged death or deterioration in the state of the health of any person.
|