| Model Number |
U225 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) recorded a fault code 1003 indicative of a battery voltage too low for the projected remaining capacity.Subsequently, this crt-p was explanted and replaced.This device was received for investigation.No additional adverse patient effects were reported.
|
| |
|