A patient underwent a percutaneous transluminal angioplasty (pta) procedure using an atlas balloon catheter.During the procedure, it was reported that the balloon failed to deflate within the subclavian vein.It was further reported that the hcp was unable to withdraw the balloon through the sheath.To facilitate safe removal, the balloon was cut at the tip and subsequently retrieved.There was no reported patient injury.
|
|
H11: this supplemental report is submitted to correct an error in the previously submitted reports.The initial mdr captured an incorrect manufacturing location (g1).The correct manufacturing location for this device is bard reynosa s.A.De c.V.No new event information is being reported in this supplement.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.However, photos were provided for review.The investigation of the reported event is currently underway.G1, g3
d2b (dqy; lit)
section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|