| Catalog Number |
15-533 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "exchange with no complaint against device".Later healthcare professional reported "intracapsular rupture".This record is for the right side.Device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "exchange with no complaint against device".Later healthcare professional reported "intracapsular rupture".This record is for the right side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional and/or corrected data: d9, h3, h6.Device evaluation:
per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
rupture: observed an opening through microscopic inspection assessed as surgical damage.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis creases are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Search Alerts/Recalls
|