• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL SYSTEMS Luer-Lok; Syringe, Piston

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL SYSTEMS Luer-Lok; Syringe, Piston Back to Search Results
Catalog Number 309628
Medical Device Problem Code Contamination (1120)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2026
Type of Reportable Event Malfunction
Event or Problem Description
No additional information was provided.
 
Additional Manufacturer Narrative
(b)(4)- supplemental mdr - foreign matter it was reported that foreign particles and plastic-like inclusions were recurring within bd 1 ml syringes.Thirteen photographs of 1 ml luer-lok syringes (part number 309628) were received and evaluated.The photographs show particulates of varying color, size, and location within the syringe barrels, including white or light-colored, brown or orange, and dark-colored particulates near the stopper and luer-end regions.The particulates appear distinct from the syringe components, stopper assemblies, and printed graduation markings, consistent with the presence of foreign matter.Additionally, three photographs show visible air bubbles within the syringe barrels near the luer end.The observed foreign matter and air bubbles are considered non-conforming per product specification, with the potential root cause associated with the assembly process.A review of the device history records (dhrs) was also completed for material number 309628, lots 3128967, 4285657, 5098930, 5121479, 5353614, and 5162348.All required visual inspections were performed, with no quality notifications identified related to the reported complaint defects.All batches met the applicable inspection control plan requirements, were accepted, and were subsequently approved for shipment.Complaints associated with this device and reported conditions will continue to be tracked and trended, with the data monitored monthly and routinely reviewed by the business team to identify any emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Luer-Lok
Common Device Name
Syringe, Piston
Manufacturer (Section D)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer (Section G)
BECTON DICKINSON MEDICAL SYSTEMS
route 7 and grace way
canaan CT 06018
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
+1(847)393-5694
MDR Report Key50169556
Report Number1213809-2026-00405
Device Sequence Number9886139
Product Code FMF
UDI-Device Identifier00382903096282
UDI-Public(01)00382903096282
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K941562
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number309628
Device Lot Number5353614
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2026
Supplement Date Received by Manufacturer09/22/2026
Initial Report FDA Received Date08/07/2026
Supplement Report FDA Received Date09/29/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-