It was reported to gore that, on (b)(6) 2026, a gore® excluder® aaa endoprosthesis was used for an endovascular aneurysm repair (evar).During preparation for introduction of the iliac component on the ipsilateral side, the 18 fr gore® dryseal flex introducer sheath was re-advanced and encountered some resistance.The clinician pushed the sheath past the resistance, causing proximal migration of the gore® excluder® aaa endoprosthesis trunk, resulting in coverage of the renal arteries.A chimney procedure was subsequently performed without complications and blood flow to both renal arteries was maintained.The patient tolerated the procedure.
|
|
As part of our routine quality procedures, each batch of devices undergoes comprehensive quality control testing and inspections prior to release for distribution.Gore reviewed the manufacturing records associated with the reported lot number and verified that all release criteria had been met.Neither clinical images enabling direct assessment of product performance nor the product itself, which remains implanted, were available for evaluation.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|