| Model Number |
MC2VR01 |
| Medical Device Problem Codes |
Failure to Capture (1081); Capturing Problem (2891); High Capture Threshold (3266)
|
| Health Effect - Clinical Codes |
Bradycardia (1751); Dizziness (2194); Presyncope (4410)
|
| Date of Event |
07/14/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that approximately two weeks after implant, the leadless implantable pulse generator (ipg) exhibited loss of capture resulting in high and rising thresholds.Patient experienced bradycardia, dizziness and near syncope as a result.The leadless ipg was programmed off and replaced.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|