Datascope Corp. SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet); System, Balloon, Intra-Aortic And Control Back to Search Results
Model Number 0684-00-0576-01
Medical Device Problem Codes Leak/Splash (1354); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Correction field: d10,g2, g7 updated fields: d1brand name, d3, d4 expiry date, version or model number, catalog number and udi number,e2,e3, b4, d9, g3, g4 (pma 510k), g6, h2, h6 (problem code, type of investigation, investigation findings, investigation conclusions, component code), h11 please revert the a1(patient id) and d8(device serviced by third party) for balloon iab d8 (device serviced by third party) should be blank the iab was returned with the membrane completely unfolded and blood on the exterior of the catheter and between the catheter and the sheath.The extender and pressure tubing were also returned.Two kinks were observed on the catheter tubing approximately 46.2cm and 76.2cm from iab tip.An underwater leak test of the balloon, catheter, y fitting, extracorporeal, extender and pressure tubing was performed and no leaks were detected.The iab was placed on the cardiosave pump and the iab fully inflated and deflated.The iab then pumped for two hours which represents one complete autofill cycle.The iab pumped normally and no alarm sounded from the pump.The reported problem cannot be confirmed by the evaluation.The failure mode is addressed in the risk file and is operating within its risk profile.The ifu addresses the reported failure.The investigation does not indicate that the device was inadvertently released as non conforming or an adulterated product or was a counterfeit.The complaint history review did not identify an adverse trend.Based on the rational provided above, no escalation to the capa process is required.
 
Event or Problem Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION PLUS 8Fr. 50cc IAB & Accessories (w/Pressure Tubes, No Stylet)
Common Device Name
System, Balloon, Intra-Aortic And Control
Manufacturer (Section D)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer (Section G)
Datascope Corp.
15 law drive
fairfield NJ 07004
Manufacturer Contact
ewelina chorlowska
15 law drive
fairfield, NJ 07004
MDR Report Key50171487
Report Number2248146-2026-0001364
Device Sequence Number12829658
Product Code DSP
UDI-Device Identifier10607567108605
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K133074
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0684-00-0576-01
Device Catalogue Number0684-00-0576-01
Device Lot Number3000508987
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/07/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2026
Supplement Date Received by Manufacturer09/22/2026
Initial Report FDA Received Date08/07/2026
Supplement Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/21/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARDIOSAVE IABP
Patient SexUnknown
-
-