MAUDE Adverse Event Report: Abbott Medical (CRM-Sylmar) Quartet Lead Double Bend, 86 cm; Drug Eluting Permanent Left Ventricular (Lv) Pacemaker Electrode

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Abbott Medical (CRM-Sylmar) Quartet Lead Double Bend, 86 cm; Drug Eluting Permanent Left Ventricular (Lv) Pacemaker Electrode Back to Search Results
Model Number 1457Q/86
Medical Device Problem Codes Failure to Capture (1081); Migration or Expulsion of Device (1395)
Health Effect - Clinical Code Dyspnea (1816)
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient presented to the emergency room experiencing worsened shortness of breath.Device interrogation revealed that the left ventricular (lv) lead failed to capture at high output.Fluoroscopy performed during the lead repositioning procedure confirmed that the lv lead had migrated into the right atrium, due to poor patient anatomy.The lv lead was explanted and replaced.There were no patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Quartet Lead Double Bend, 86 cm
Common Device Name
Drug Eluting Permanent Left Ventricular (Lv) Pacemaker Electrode
Manufacturer (Section D)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
Abbott Medical (CRM-Sylmar)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key50171686
Report Number2017865-2026-16584
Device Sequence Number8242877
Product Code OJX
UDI-Device Identifier05414734510189
UDI-Public(01)05414734510189(10)A000150229(17)261031
Combination Product (Y/N)Y
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1457Q/86
Device Lot NumberA000150229
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/28/2026
Initial Report FDA Received Date08/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/10/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
Quadra Allure MP; UltiPace; UltiPace
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient SexFemale
-
-