Medtronic Mexico S. De R.L. De CV VitalFlow Set; BLOOD PUMP FOR ECMO, LONG-TERM (> 6 HOURS) USE
|
Back to Search Results |
|
| Model Number |
58112 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Medtronic received information that during use of a vital flow tubing set, it was reported that the device had a flow/ clot issues.The biomed believes the device sucked a clot from the left ventricle of the patient into the pump head.Flow stopped and a hand crank was used for roughly 10 mins, and this provided some flow.Customer placed the cone back to the motor drive and was able to get some flow the customer believed the clot in the left ventricle of the heart became dislodged and migrated into the motor drive.The patient was previously on heparin, the issue did not worsen over time.The rpm at time of event was 3800, the flow at time of event- 0.33lpm, and the only blood loss was the blood in the clotted off circuit.Anticoagulation used heparin 100 units/hr, not acts were collected, aptt 40-50.Plasmalyte was used to prime the circuit.The location of this issue was at the cone and pre ox side of vital flow membrane oxygenator. the device was replaced to complete the procedure.There was no adverse patient effect associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction b5: medtronic received information that during use of a vitalflow tubing set, it was reported that the device had a flow/clot issue.The physician believes the device sucked a clot from the left ventricle of the patient into the pump head.Flow stopped and a hand crank was used for roughly 10 minutes, and this provided some flow.The customer placed the cone back to the motor drive and was able to get some flow.The anticoagulation used was heparin 100 units/hr, no acts were collected, aptt 40-50.The location of this issue was at the cone and pre-oxygenator side of the vitalflow membrane oxygenator. correction b2(outcome attributed to adverse event), d2 (product code), d2.1 (common device name), h6.3 (eval code result (fdr/annex c)), additional codes (imf): these fields have been updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation summary: visual inspection shows evidence of clots inside the device and around the impeller upper pivot.Additional visual inspection shows evidence of damage around the impeller upper pivot due to the clot not allowing the blood to lubricate the pivot.The device was tested per specification. the device was run on a bio console from 0-4000 rpm¿s, using di water in the circuit. with no clots inside the device there was a flow of 7 lpm observed.Reason for the return was confirmed. medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|