|
Additional, changed, and/or corrected data: b6, d9, e3, h3, h6.Lab results:
per the investigation procedure, the device is analyzed through visual microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
rupture: observed an opening through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis creases and stress marks are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|