• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Intervascular SAS Intergard Woven Aortic Arch; Prosthesis, Vascular Graft, Of 6mm And Greater Diameter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Intervascular SAS Intergard Woven Aortic Arch; Prosthesis, Vascular Graft, Of 6mm And Greater Diameter Back to Search Results
Model Number HEWAA3010080810/1
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2026
Type of Reportable Event Malfunction
Event or Problem Description
Complaint #(b)(4).
 
Additional Manufacturer Narrative
(10/213) the returned product was visually inspected by the quality assurance (qa) supervisor.The analysis of the product confirms the presence of the yellow stains on the graft surface.Such spots are a known occurrence, and it can be confirmed that they are caused by collagen.But this is a known phenomenon that doesn't affect the product's performance or safety if no additional thickness is visible on the product, which is the case for the product.The product therefore complies with the specifications.(67) based on the investigation findings, it was concluded that the involved product is conform to the product specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Intergard Woven Aortic Arch
Common Device Name
Prosthesis, Vascular Graft, Of 6mm And Greater Diameter
Manufacturer (Section D)
Intervascular SAS
zone industrielle athelia i
la ciotat 13705
FR  13705
Manufacturer (Section G)
Intervascular SAS
zone industrielle athelia i
la ciotat 13705
FR   13705
Manufacturer Contact
clemence vaneenoge
zone industrielle athelia i
la ciotat 13705
FR   13705
+033()442084646
MDR Report Key50173515
Report Number1640201-2026-0000037
Device Sequence Number9880594
Product Code DSY
UDI-Device Identifier00384401010760
UDI-Public(01)00384401010760
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K013651
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/30/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHEWAA3010080810/1
Device Catalogue NumberHEWAA3010080810/1
Device Lot Number24L14
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer08/25/2026
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/17/2026
Supplement Date Received by Manufacturer09/30/2026
Initial Report FDA Received Date08/07/2026
Supplement Report FDA Received Date09/30/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/14/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-