| Model Number |
CCWTT3 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Visual Disturbances (2140)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A non-healthcare professional reported that following an intraocular lens (iol) implantation procedure, the patient had experienced blurred vision, halos and unhappy with vision.The lens was rotated and repositioned by the surgeon approximately three months after the surgery.Additional information has been requested but is not available at the time of this report.There are two medical device reports associated with this patient.This report is associated with the left eye.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for analysis.Complaint history and product history records were reviewed, and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
|
| |
|