| Model Number |
SGC0701 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 4+ functional mitral regurgitation (mr) and restricted posterior leaflet for a mitraclip procedure.During the preparation, it was noticed that hemostatic valve on the back of the guide catheter was not holding column when dilator was inserted.Device was replaced and continued the procedure.All steps were performed as per ifu.The final mr was grade 1-2.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, the cause of the reported loss of fluid column was unable to be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|