Hologic, Inc. EVIVA_0913-20; INSTRUMENT, BIOPSY Back to Search Results
Model Number EVIVA_0913-20
Medical Device Problem Code Failure to Cut (2587)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported by the user site that during a breast biopsy procedure the device wasn't cutting.Additional information obtained from the user site confirmed that after targeting, firing, and attempting to obtain a sample, "the doctor identified that the needle was not cutting when using the foot pedal." the customer also confirmed "the patient was rescheduled and returned the next week to complete the procedure." additionally, the customer confirmed the needle in question was from lot e26de0ru.No further information is availbe at this time.
 
Additional Manufacturer Narrative
Device history record (dhr) review was unable to be conducted for the device as the reported lot number (e26de0ru) did not populate in the hologic system.A follow-up report will be submitted in the event additional information is received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIVA_0913-20
Common Device Name
INSTRUMENT, BIOPSY
Manufacturer (Section D)
Hologic, Inc.
250 campus drive
marlborough MA 01752
Manufacturer (Section G)
Hologic, Inc.
250 campus drive
marlborough MA 01752
Manufacturer Contact
sarah lawlor
250 campus drive
marlborough, MA 01752
MDR Report Key50175564
Report Number1222780-2026-00485
Device Sequence Number13231741
Product Code KNW
UDI-Device Identifier15420045504974
UDI-Public15420045504974
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K180233
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberEVIVA_0913-20
Device Catalogue NumberEVIVA_0913-20
Device Lot NumberE26DE0RU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2026
Initial Report FDA Received Date08/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-