Neither the device nor photographic evidence was provided.Therefore, the reported event cannot be verified.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.A two-year lot history review shows a total of one (1) device for this lot number and failure mode.A two-year review of complaint history revealed there has been a total of 21 reports, regarding 21 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised that if the hook or needle bends during use, immediately discontinue use and discard.There is an increased risk of hook or needle breakage, and unintentional patient injury may result.Avoid lateral stresses to the instruments, or device function may be compromised.There is an increased risk of hook or needle breakage, and unintentional patient injury may result.Do not use if parts are broken, cracked, or worn, or device function may be compromised.There is an increased risk of hook or needle breakage, and unintentional patient injury may result.We will continue to monitor per regulatory requirements to help ensure patient safety.
|