| Model Number |
AFR-00001 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Bradycardia (1751); Hemorrhage/Blood Loss/Bleeding (1888); Heart Block (4444)
|
| Date of Event |
07/07/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was later reported that a "temperature spread too high" error code was received during the procedure that did not allow delivery of ablations.Several troubleshooting techniques were attempted, including replacement of the catheter extension, but the issue was not resolved.The was catheter replaced, and the issue was resolved.The case was completed.
|
| |
|
Event or Problem Description
|
|
It was later reported that sinus bradycardia with 1 degree av block was observed prior to and after ablation procedure.
|
| |
|
Event or Problem Description
|
|
It was reported that during a repeat cardiac ablation procedure, sinus bradycardia with first-degree atrioventricular block was observed at the end of procedure.It was noted that patient undergone pulmonary vein isolation (pvi) procedure. it was unknown whether sinus bradycardia with first-degree atrioventricular block were resolved.It was further reported that post-procedure groin bleeding occurred, and the patient was monitored overnight.The following morning, the groin suture was removed, with no further bleeding or hematoma reported.No changes were made to the medication regimen, and no acute events occurred overnight.The groin bleeding resolved, and the patient was discharged.The case outcome is unknown.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
Continuation of d10: product id: afr-00001 product type: catheter.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|