| Model Number |
25AGN-751 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
Aortic Valve Stenosis (1717)
|
| Date of Event |
07/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on(b)(6) 2026, a 25mm sjm regent heart valve was chosen for a procedure.After implantation, an increased pressure gradient was noted.A tear in the valve ring was confirmed by visual observation.The valve got explanted and a 23mm sjm regent heart valve was implanted.The patient was reported stable.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of explant due to aortic valve stenosis was reported.Field indicated that it was found to a tear in the valve ring was confirmed by visual observation.A returned device could not be performed as the device was not returned to abbott.Abbott also requested for image of the device/issue and procedural conditions.The device history record was also reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.Based on received information, the cause of reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|