| Model Number |
NVRO-29 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Heart Block (4444)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a 29mm navitor vision valve was chosen for a transcatheter aortic valve replacement (tavr) procedure.The length of the membranous septum was 6mm.There was no calcification present extending beneath the aortic annular plane in the interventricular septum.The final implant depth of the valve was 3mm.Post procedure, the patient needed a permanent pacemaker for complete heart block (chb) after tavr.The patient required prolonged hospital stay for monitoring of rhythm and the patient was reported discharged.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a 29mm navitor vision valve was chosen for a transcatheter aortic valve replacement (tavr) procedure.The length of the membranous septum was 6mm.There was no calcification present extending beneath the aortic annular plane in the interventricular septum.The final implant depth of the valve was 3mm.Post procedure, the patient needed a permanent pacemaker for complete heart block (chb) after tavr.The patient required prolonged hospital stay for monitoring of rhythm and the patient was reported discharged.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of heart block was reported intra-procedure.The reported surgical intervention was result of case specific circumstance as permanent pacemaker was implanted.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.Based on the available information the cause for the reported heart block cannot be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Additional Manufacturer Narrative
|
|
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|