• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. ENDOSCOPIC CO2 REGULATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. ENDOSCOPIC CO2 REGULATION UNIT Back to Search Results
Model Number UCR
Medical Device Problem Code Increase in Pressure (1491)
Health Effect - Clinical Code Bowel Perforation (2668)
Date of Event 07/10/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after a colonoscopy procedure, the patient had a 2cm serosal tears at the cecum due to overinflation of carbon dioxide (co2) using this co2 regulation unit.An upright chest x-ray revealed a free intraperitoneal air, as per the physician.It was not a through and through but a damage to the wall of cecum due to overstretching.The patient was brought back to the operating room and a exploratory laparotomy, ileocecectomy were performed.The patient was then transferred to the intensive care unit (icu) and is currently stable.There were no further reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOSCOPIC CO2 REGULATION UNIT
Common Device Name
CO2 REGULATION UNIT
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
MDR Report Key50178247
Report Number2429304-2026-00108
Device Sequence Number21303935
Product Code HIF
UDI-Device Identifier04953170239113
UDI-Public04953170239113
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 08/07/2026
Report Date (Section F) 08/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberUCR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date07/16/2026
Device Age8 YR
Event Location Hospital
Date Report to Manufacturer07/16/2026
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/07/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
Date Report Sent to FDA08/07/2026
-
-