| Model Number |
FNAV-DS-LG |
| Medical Device Problem Code |
Retraction Problem (1536)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a 29 navitor vision valve was selected for implant using a large flexnav delivery system (ds).During the procedure.While attempting to recapture on the second attempt, the valve was unable to be recaptured into the ds.It was also reported that micro-adjustment wheel was used while the ds was in the descending aorta.A new 29mm, navitor vision valve was replaced using a new large flexnav ds resulting in a successful implantation.The patient remained hemodynamically stable throughout the procedure and there was no clinically significant delay reported.The patient was reportedly discharged.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|