| Model Number |
3CX*FX25RWC |
| Medical Device Problem Code |
Infusion or Flow Problem (2964)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Terumo has received the device for evaluation; however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.For this reason, terumo references evaluation conclusion code 11.H3: evaluation is in progress, but not yet concluded. .
|
| |
|
Event or Problem Description
|
|
The user facility reported to terumo cardiovascular that prior to cardiopulmonary bypass, during prime, they noticed a rapid de-priming of the venous line after pump flow was stopped.*no patient involvement.*product was changed out.*procedure completed successfully.
|
| |
|