| Model Number |
DKPI-11073-001 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
An initial event notification was received by sequel med tech, llc, on (b)(6) 2026 and forwarded to millyard advanced medical products, llc, on the same day.The user reported that the pigtail tubing detached from their twiist cassette when they rolled over in their sleep.The user changed their cassette and resumed therapy without issue.The user remains ongoing on their twiist pump.
|
| |
|
Additional Manufacturer Narrative
|
|
Cassette (b)(4) was returned for investigation with the cassette pigtail tubing detached.Inspection of the cassette confirmed damage to the tubing, with evidence indicating that there was excessive pulling force that caused the tubing to stretch, fray, and break.The user remains ongoing on their twist pump.The twist automated insulin delivery system user guide instructs users to have a backup therapy available at all times.No additional information relevant to the reported event has been made available to millyard advanced medical products, llc, for further investigation.
|
| |
|