Model Number IntelliVue MP5
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2026
Type of Reportable Event Malfunction
Event or Problem Description
Philips received a complaint on an intellivue mp5 indicating that the blood pressure (bp) tubing was kinked causing device to not take accurate readings.The device was not in clinical use at the time of the event.No adverse event was reported.
 
Additional Manufacturer Narrative
Analysis was performed by onsite biomedical personnel which included inspection of the system.The results of the analysis were that the hose was kinked.Based on the results of the analysis, the product was confirmed to be operating per specifications, and the cause of the reported problem was a kinked hose.The issue was resolved by removing the kink from the hose.Device is working as intended, unit returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IntelliVue MP5
Common Device Name
Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Manufacturer (Section D)
Philips Medizin Systeme Böblingen GmbH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
Philips Medizin Systeme Böblingen GmbH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
+49(703)14632031
MDR Report Key50181047
Report Number9610816-2026-101613
Device Sequence Number8271122
Product Code MHX
UDI-Device Identifier00884838000230
UDI-Public(01)00884838000230
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K150310
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 08/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIntelliVue MP5
Device Catalogue Number865024
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2026
Initial Report FDA Received Date08/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/12/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1