| Model Number |
IntelliVue MP5 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Philips received a complaint on an intellivue mp5 indicating that the blood pressure (bp) tubing was kinked causing device to not take accurate readings.The device was not in clinical use at the time of the event.No adverse event was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Analysis was performed by onsite biomedical personnel which included inspection of the system.The results of the analysis were that the hose was kinked.Based on the results of the analysis, the product was confirmed to be operating per specifications, and the cause of the reported problem was a kinked hose.The issue was resolved by removing the kink from the hose.Device is working as intended, unit returned to service.
|
| |
|