| Model Number |
CDVRA500T |
| Medical Device Problem Code |
Pacing Problem (1439)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient's implantable cardioverter defibrillator (icd) exhibited low pacing output.The physician elected to explant and replace the icd.The patient condition was not known.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported complaint of low pacing output issues was not confirmed.The device was received in normal range of operation.Analysis of device image was performed and no anomalies were noted.Further electrical testing was performed, including impedance and output testing, and no issues were noted.Longevity assessment found the device was operating above elective replacement indicator (eri) level and within expected longevity.
|
| |
|