During an atrial fibrillation procedure, st elevation was noted during a guidewire exchange in the fast-cath sheath.Following difficult femoral venous access and transseptal puncture, the physician opted to switch to a stiffer guidewire to help advancement of the agilis sheath through the kinked femoral vein.The fast-cath sheath was in the left atrium, and the guidewire was located in the left superior pulmonary vein.While withdrawing the guidewire, the patient developed st elevation on the ecg.It was noted that the guidewire was not withdrawn with the dilator, which is alleged to have caused air bubbles that presumably got into the coronaries.The patient became hemodynamically unstable, pulse was lost, asystole on ecg and decrease in blood pressure.Akrinor was administered for hemodynamic support.Cpr was performed for a few seconds.Heartbeat was obtained with only a rate of approximately 30 beats per minute.Emergent ventricular pacing was started and performed by an inquiry catheter which was placed in the rv.The st elevation was noted to gradually resolve over the next few minutes.The patient's hemodynamic condition slowly stabilized and ventricular pacing was able to be discontinued.The procedure was cancelled and the patient was transferred out of the operating room in stable condition.
|