• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: Nihon Kohden Corporation ZM-531PA; Telemetry Transmitter

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

Nihon Kohden Corporation ZM-531PA; Telemetry Transmitter Back to Search Results
Model Number ZM-531PA
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/15/2026
Type of Reportable Event Malfunction
Event or Problem Description
The biomedical engineer (biomed) reported that the zm transmitter would intermittently drop its connection with the cns.No patient harm was reported.
 
Additional Manufacturer Narrative
The biomedical engineer (biomed) reported that the zm transmitter would intermittently drop its connection with the cns.No patient harm was reported.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.The following field contains no information (ni), as attempts to obtain the information were made, but not provided: d10: "[remove if all needed information was provided, including all model and serial numbers.]" attempt # 1: (b)(6) 2026 emailed the customer via microsoft outlook for the information in the ni list above: no reply was received.Attempt # 2: (b)(6) 2026 emailed the customer via microsoft outlook for the information in the ni list above: no reply was received.Attempt # 3: (b)(6) 2026 emailed the customer via microsoft outlook for the information in the ni list above: no reply was received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-531PA
Common Device Name
Telemetry Transmitter
Manufacturer (Section D)
Nihon Kohden Corporation
attn: shama mooman
1-31-4 nishiochia
shinjuku-ku, tokyo 161-8 560
JA  161-8560
Manufacturer (Section G)
Nihon Kohden Tomioka Corporation
attn: shama mooman
1-1 tajino
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
MDR Report Key50181817
Report Number8030229-2026-06962
Device Sequence Number9886071
Product Code DRT
UDI-Device Identifier04931921115107
UDI-Public(01)04931921115107
Combination Product (Y/N)N
Initial Reporter StateID
Initial Reporter CountryUS
PMA/510(K) Number
K043517
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 08/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZM-531PA
Device Catalogue NumberZM-531PA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/15/2026
Initial Report FDA Received Date08/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/15/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-