| Model Number |
1948/58 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
08/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This report is to advise of an event observed during analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
No complaint was received with the return of the device.Failure event observed during analysis.All tines were intact, and no anomalies were noted.Visual inspection of the lead found an external insulation abrasion exposing the outer coil consistent with friction to the device can, located in the proximal region of the lead.All other damages were found consistent with procedural damage.
|
| |
|