| Model Number |
CYF-VH |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: date of event is unknown.Additional information will be provided if available.The device was returned to olympus for inspection.Based on the results of the investigation, a definitive root cause of the missing/detached objective lens adhesive could not be determined, however, the issue was likely the result of component failure due to degradation, handling and or external factors.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The cysto-nephro videoscope was returned to the service center for a separate issue.During evaluation of the device, the objective lens adhesive was found to be missing/detached, likely due to degradation, and was repaired.There was no patient involvement, and no harm was reported as a result of the event.
|
| |
|