| Model Number |
1610010 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Low Oxygen Saturation (2477)
|
| Date of Event |
07/06/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Rt was called to bedside due to nursing stating that ventilator was spontaneously shutting off.Rt came to bedside and nursing stated it happened while suctioning.Rt suctioned patient and all flow from the ventilator stopped completely and patient o2 saturations dropped to low 80's and required support from ambu bag to improve saturations.Rt replaced flow sensor and exhalation valve and still the machine shut off all flow after suctioning despite the patient spontaneously breathing.
|
| |
|