Product analysis as found condition (condition of returned device): the concerto implant coil as found images are unavailable.The concerto implant device was returned within the introducer sheath.Visual inspection/damage location details: the concerto coil pushwire was found to be bent within introducer sheath at ~ 87.2cm and kinked within introducer sheath at ~164.0cm from proximal end.The implant coil was found to be already detached and not returned.All other subassemblies appeared to be normal, and no other anomalies were observed.Testing/analysis (including sem reports): the concerto coil could not be used for resistance testing with an in-house micro catheter due to its damaged condition.Conclusion: based on the device analysis and the reported information, the customers report of ¿coil resistance/stuck in cath¿ was unable to be confirmed.The model or lot number for the unknown catheter were not provided; therefore, compatibility could not be assessed.The root cause could not be determined.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|