| Model Number |
BF-1T150 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer returned the broncho videoscope to the service center for a separate issue.During olympus¿ device analysis it was indicated that the broncho videoscope's bending section cover adhesive was detached near the distal end.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection when the reportable malfunction was observed.Based on the results of the investigation, the observed issue was attributed to degradation problems linked to the device, however, a definitive root cause could not be established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|