Model Number NV-3-4-HELIX
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/04/2026
Type of Reportable Event Malfunction
Event or Problem Description
Medtronic received a report that the concerto coil could not be advanced.The procedure was completed by using a new product, and the causes or contributing factors for the coil not advancing were not identified.The reported device and any accessory devices were prepared as indicated in the instructions for use (ifu).No patient symptoms or complications were associated with this event.
 
Additional Manufacturer Narrative
Product analysis ¿ as found condition: the concerto device was returned for analysis within a shipping box, within a sealed plastic biohazard pouch, within an opened concerto outer carton and within an opened concerto inner pouch.The unknown micro catheter used in the event was not returned for analysis.¿ visual inspection/damage location details: the concerto pusher was found broken distal to the break indicator with the release wire also broken.The proximal segment was not returned for analysis.The concerto implant coil was found still attached to the pusher and found damaged.¿ testing/analysis: the concerto coil could not be used for resistance testing with an in-house micro catheter due to the damaged condition.¿ conclusion: based on the device analysis and reported information, the customer¿s report of ¿coil resistance/stuck in catheter¿ was confirmed as the damages found is consistent with resistance.The likely cause could not be determined.The returned concerto implant coil was found damaged.The customer did not report any damage during inspection/preparation per ifu; therefore, it is possible the damage occurred due to advancing against the reported resistance.The root cause could not be determined.As the unknown micro catheter used in the event was not returned for analysis, any contribution of the micro catheter towards the resistance could not be determined.As the model and lot numbers were not reported, compatibility could not be assessed.The pusher was found broken and the release wire was found broken.The likely cause could not be determined.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONCERTO NYLON
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
Medtronic Mexico S. De R.L. De CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX  22210
Manufacturer (Section G)
Medtronic Mexico S. De R.L. De CV
av. paseo del cucapah #10510
tijuana,bc 22210
MX   22210
Manufacturer Contact
justin ellis
9775 toledo way
irvine, CA 92618
+1(763)526-5677
MDR Report Key50183554
Report Number9617601-2026-05016
Device Sequence Number10169205
Product Code KRD
UDI-Device Identifier00847536044890
UDI-Public(01)00847536044890(17)281007(10)232434382
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K090046
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNV-3-4-HELIX
Device Catalogue NumberNV-3-4-HELIX
Device Lot Number232434382
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/02/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/04/2026
Initial Report FDA Received Date08/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/07/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-