| Model Number |
15 |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/21/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer contacted stryker to report that their device did not shock.In this state, defibrillation therapy may be delayed or prevented from being delivered to the patient if needed.This issue is patient related; however, there was no adverse patient outcome reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Stryker contacted the customer to request additional information on the patient.Stryker requests patient information necessary for regulatory compliance, strictly adhering to hipaa's privacy rule.The customer was unable to provide additional patient information.Patient fields in which information is not provided were intentionally left blank.Stryker evaluated the customer's device and was unable to duplicate the reported issue.It was observed that the device logged an event code in the device¿s memory that is related to a potential failure of the device to deliver defibrillation energy.The main keypad was replaced to resolve the issue.After other unrelated repairs were completed, proper device operation was observed through functional and performance testing, and the device was returned to the customer for use.
|
| |
|