HistoSonics, Inc. Edison; Focused Ultrasound System For Non-Thermal, Mechanical Tissue Ablation
|
Back to Search Results |
|
| Catalog Number |
EDNTC1001 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Renal Impairment (4499)
|
| Date of Event |
04/04/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As part of the boombox post-market clinical study, on (b)(6) 2025, a 77 year old female patient with a history of hepatocellular carcinoma received histotripsy treatment to two lesions in liver segments vi and vii for a total planned treatment volume (ptv) of approximately 43.4 cc.As per treating facility standard of care, the patient was held overnight for observation and discharged the following day.Later that day, the patient developed red-tinged sputum with clots.The patient contacted the treating physician and was advised to present to an emergency department for evaluation.On (b)(6) the patient was admitted to a local hospital with presumptive pneumonia and acute kidney injury although creatinine values were not available (baseline or elevated).The patient remained hospitalized until transfer to the treating facility was initiated on (b)(6).The patient was evaluated by the treating physician on (b)(6).After review of the outside hospital records and imaging, the treating physician documented that the patient did not appear to have pneumonia and that the pulmonary findings were more likely atelectasis related to single-lung ventilation during the histotripsy procedure.Laboratory testing on (b)(6)showed improved renal function, with creatinine decreased to 1.78 mg/dl from a reported value greater than 3 mg/dl.The acute kidney injury was documented as significantly improved.No dialysis or renal replacement therapy was reported.The patient was discharged home on (b)(6) with follow-up arranged in clinic.
|
| |
|
Additional Manufacturer Narrative
|
|
No device malfunctions or other noteworthy events occurred during the histotripsy procedure.The post-procedural acute kidney injury is consistent with the labeled risk associated with large treatment volumes and multiple treatment sessions.Histosonics has incorporated the following warning into its labeling: "consider patient-specific risk factors for acute kidney injury (aki) and/or renal failure based on clinical history and procedural scenarios that may increase stress on kidney function.Factors to consider include, but are not limited to, hydration status, planned treatment volume, number of treatment sessions, use of nephrotoxic agents including nsaids, and expected use of imaging contrast agents.The risk of aki increases as total treatment volume increases.In patients with compromised pre-procedure renal function, aki may progress to acute renal failure.Renal risk should be evaluated before treatment, and renal function should be monitored for signs of aki before, during, and after treatment.".
|
| |
|
Search Alerts/Recalls
|
|
|