• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN AG JUAREZ (MMJ) Shiley; TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN AG JUAREZ (MMJ) Shiley; TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number 18875
Medical Device Problem Code Gas/Air Leak (2946)
Health Effect - Clinical Code Unintended Extubation (4564)
Date of Event 05/14/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after the patient developed weak breathing due to intracerebral hemorrhage, endotracheal intubation was ordered by the physician and the intubation was successful.After intubation, the patient was connected to a ventilator for assisted breathing, but the ventilator issued a continuous low-pressure alarm.The alarm persisted even after repeated adjustments, the cuff pressure of the tube was checked and found to be slightly low, then the cuff was inflated.After inflation, the ventilator alarm disappeared then about 2 minutes later, the ventilator again issued a low-pressure alarm, which happened repeatedly.The tube was replaced, and when the removed tube was inspected, it inflated smoothly and the cuff was full.After water was injected into the cuff, continuous water jets were observed from the cuff wall, confirming cuff leakage.The event delayed the patient¿s treatment but no apparent harm.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Shiley
Common Device Name
TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX  32575
Manufacturer (Section G)
COVIDIEN AG JUAREZ (MMJ)
1181 hennequen ave.
juarez,chi 32575
MX   32575
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112--439
+1(800)646-4633
MDR Report Key50184595
Report Number9681384-2026-00791
Device Sequence Number13286844
Product Code BTR
UDI-Device Identifier40884521032539
UDI-Public(01)40884521032539(17)300626(10)25F1288JZX
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K090352
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Type of Report Initial
Report Date (Section B) 08/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number18875
Device Catalogue Number18875
Device Lot Number25F1288JZX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/30/2026
Initial Report FDA Received Date08/08/2026
Date Device Manufactured06/26/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age74 YR
Patient SexUnknown
-
-