It was reported that after the patient developed weak breathing due to intracerebral hemorrhage, endotracheal intubation was ordered by the physician and the intubation was successful.After intubation, the patient was connected to a ventilator for assisted breathing, but the ventilator issued a continuous low-pressure alarm.The alarm persisted even after repeated adjustments, the cuff pressure of the tube was checked and found to be slightly low, then the cuff was inflated.After inflation, the ventilator alarm disappeared then about 2 minutes later, the ventilator again issued a low-pressure alarm, which happened repeatedly.The tube was replaced, and when the removed tube was inspected, it inflated smoothly and the cuff was full.After water was injected into the cuff, continuous water jets were observed from the cuff wall, confirming cuff leakage.The event delayed the patient¿s treatment but no apparent harm.
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|