| Model Number |
380723-12 |
| Medical Device Problem Code |
Connection Problem (2900)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to starting in a da vinci-assisted surgical procedure, no ports placed, the customer identified vision side cart (vsc) did not progress self-test and surgeon side console (ssc) displayed message not connected to vision side cart (vsc).Technical support engineer (tse) observed recoverable error message/fault 31016, pointing at ssc being at different software level than vsc.The customer elected to swap ssc available on site to continue with the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
An intuitive field service engineer (fse) was dispatched to the customer site to further investigate the reported event.Based on the field evaluation, the fse updated software on surgeon side console (ssc).Service performed to correct reported problem.No part replacement required.System inspected, tested and verified as ready for use.
|
| |
|