MAUDE Adverse Event Report: AMEDA INC. PURELY YOURS; ELECTRIC BREAST PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMEDA INC. PURELY YOURS; ELECTRIC BREAST PUMP Back to Search Results
Model Number 24502082
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/13/2015
Event Type  malfunction  
Manufacturer Narrative
The returned product was tested per ameda engineering protocol to determine whether the allegation of leaking fluid could be confirmed.The returned product was assessed for indications of malfunction or thermal event.The returned product was assessed for functionality and met functional specifications.Dark grey substance, e.G battery acid, was observed inside the battery compartment.Internally, no signs of foreign substance, burning, melting or charring were observed.The returned batteries were observably damaged.Additional testing confirmed that batteries may leak when the upper middle battery is placed incorrectly, with the positive and negative end reversed.
 
Event Description
Customer contacted ameda, inc.On (b)(6) 2015 to report pumping on battery power with her purely yours breast pump and a battery exploded inside the battery compartment immediately after completing her pumping session.She states hearing a hissing sound from the compartment prior to the event occurring.Battery leaked black fluid inside the compartment but she denies any burn, injury or property damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS
Type of Device
ELECTRIC BREAST PUMP
Manufacturer (Section D)
AMEDA INC.
buffalo grove IL
Manufacturer Contact
485 e half day rd ste 320
buffalo grove, IL 60089
8479642620
MDR Report Key5018611
MDR Text Key23874041
Report Number3009974348-2015-00144
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Nurse
Remedial Action Replace
Report Date 07/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/07/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number24502082
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer07/17/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/13/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/11/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-