| Model Number |
NIMEID080 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Codes |
Unspecified Tissue Injury (4559); Unintended Extubation (4564)
|
| Date of Event |
07/14/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that patient having the removal of a right hemi thyroid.The anesthetist used the nim essence tubes and spent over 30 minutes trying to intubate.The anesthetist had the nim vital on the setup electrode check screen and they said they spent all this time trying to achieve the green confirmation to proceed ticks, every slight movement would mean that the ticks would go back to the red cross.So small adjustments slightly left then 5mm slight right to get the tube to be sat in the correct position and the electrode check kept changing from green ticks back to red ticks with no room for any slight error. they also re-intubated the patient with a second tube to have the same issue that only resolved when they used the nim flex tube.There was 30 min delay in the procedure.The patient suffered soft palet injury due to repeat re-intubation.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|