| Model Number |
AIA-360 IMMUNOASSAY ANALYZER |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer reported failed american proficiency institute (api) second event proficiency testing for free triiodothyronine (ft3) on the aia-360 analyzer.Samples ch-06 and ch-10 failed, and no additional information has been provided.Tosoh quality laboratory participated in the american proficiency institute (api) second event proficiency testing for free triiodothyronine (ft3) and passed all results.A field service engineer (fse) was dispatched to address the reported event.There was no indication of any patient intervention or adverse health consequences due to the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
The cause of the reported event has not been determined.The investigation is ongoing.
|
| |
|
Search Alerts/Recalls
|