| Model Number |
380677-17 |
| Medical Device Problem Code |
Visual Prompts will not Clear (2281)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that prior to start- pre anesthesia in a da vinci-assisted surgical procedure, no ports placed, the customer identified multiple non- recoverable error messages/ faults on the system while starting the system.The customer rebooted the system using a hard power cycle and swapped and cleaned the blue fiber cable (bfc) but issue did not resolved.The customer decided to proceed laparoscopically without contacting the technical support engineer (tse).Tse contacted customer after the procedure was completed and asked to perform another power cycle, but the issue persisted.Tse checked system logs and confirmed error message/ fault pointing towards personality module surgeon console (pmsc) of the surgeon side console (ssc).
|
| |
|
Additional Manufacturer Narrative
|
|
An intuitive field service engineer (fse) was dispatched to the customer site to further investigate the reported event.Based on the field evaluation, the fse replaced personality module surgeon console (pmsc) in the surgeon side console (ssc).The system was tested and verified as ready for use.Intuitive surgical inc., (isi) did not received the product for failure analysis evaluation.
|
| |
|