Hoya corporation became aware of this event on or before july 31, 2026.Effective august 1, 2026, mdr-related activities and responsibilities were transferred to pentax medical corporation as part of a corporate restructuring.This report is being submitted by pentax medical corporation.In accordance with 21 cfr 803.3(m), because this corporate restructuring does not affect the manufacturing site, the fda registration number previously assigned for this site (#9610877) continues to be used.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the water nozzle missing.Based on the result, we concluded that it was caused due to the physical damage applied on the water nozzle.In addition, our technician confirmed that the control body corroded, the operation channel (primary) resistance, the angulation down angulation decrease, and the angulation left angulation decrease; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.As a result of confirming the detail as gfe, no response was obtained, so we cannot deny the possibility of the nozzle falls.Therefore, based on the technical report "hr-rpt-0585 (nozzle)" and/or the risk analysis results, it was evaluated to submit mdr.
|