|
Hoya corporation became aware of this event on or before july 31, 2026.Effective august 1, 2026, mdr-related activities and responsibilities were transferred to pentax medical corporation as part of a corporate restructuring.This report is being submitted by pentax medical corporation.In accordance with 21 cfr 803.3(m), because this corporate restructuring does not affect the manufacturing site, the fda registration number previously assigned for this site (#9610877) continues to be used.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb distorted image.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module with drive pcb.Based on the technical report "hr-rpt-0586(image failure)" and/or the risk analysis results, it was evaluated to submit mdr.
|