Bard Peripheral Vascular, Inc. Vaccess; PTA Balloon Dilatation Catheter Back to Search Results
Catalog Number VA80124
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/29/2026
Type of Reportable Event Malfunction
Event or Problem Description
On (b)(6) 2026, a 62 year old hispanic male patient underwent a percutaneous transluminal angioplasty (pta) procedure using a vaccess balloon catheter.During the procedure, the balloon allegedly experienced a pinhole rupture.There was no reported patient injury.
 
Additional Manufacturer Narrative
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one vaccess pta dilatation catheters were returned for evaluation.Upon visual inspection kinks were noted to catheter at the proximal end of the balloon, no anomalies were noted to the luers.During functional testing, the balloon was inflated and water was noted leaking from the proximal end of the balloon.Under microscopic examination a compound rupture was noted to the balloon.No other functional testing performed.Therefore, the investigation is confirmed for the reported balloon rupture as a compound rupture was noted to the balloon.The presence of stents at the treatment site could have caused the balloon rupture.However, the definitive root cause for the balloon rupture could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.D2b (dqy; lit).Section a through f: the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Vaccess
Common Device Name
PTA Balloon Dilatation Catheter
Manufacturer (Section D)
Bard Peripheral Vascular, Inc.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
Futurematrix Interventional
1605 enterprise street
athens 75751
Manufacturer Contact
michael gaffney
800 w. rio salado pkwy
tempe, AZ 85281
+1(480)303-2689
MDR Report Key50189804
Report Number2020394-2026-02050
Device Sequence Number17669955
Product Code LIT
UDI-Device Identifier00801741121265
UDI-Public(01)00801741121265(17)290131(10)93NL0156
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K111850
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 08/03/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVA80124
Device Lot Number93NL0156
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer05/26/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 08/03/2026
Initial Report FDA Received Date08/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age62 YR
Patient SexMale
Patient Weight125 LB
-
-