| Model Number |
ELCTRNC VENOUS OCCLDR UNIT |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/13/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Livanova deutschland received a report that electronic venous occluder (evo) displayed an error code indicating that it was not working properly during procedure.Reportedly, error acknowledgement temporarily cleared after cleaning procedure and then it reappeared.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
A1-a5 patient information was not provided.D.4.Serial number is unknown.This information will be provided in a supplemental report if made available.H.4.As the serial number is unknown, the device manufacture date could not be determined.This information will be provided in a supplemental report if made available.H11 livanova deutschland manufactures the evo clamp, the event occurred in japan.Through follow up communication it was learned that the displayed error was related to a faulty encoder (e1538).If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
| |
|